LOCATION:HOME > SERVICES >

SERVICES

10 Best Practices for Electrosurgical Unit Testi

TIME:2021-06-23 17:59 点击次数:


Tips to quickly and effectively test electrosurgical devices to ensure performance and safety

Let’s start with the basics Electrosurgical units (ESU) use a high-frequency electrical current to cut tissue and control bleeding by causing coagulation. Tissue resistance to the high-density current causes a heating effect which results in tissue destruction. Electrical current is delivered and received through cables and electrodes. The electrodes may be activated by either a hand piece switch or a footswitch. The ESU may use a monopolar or a bipolar mode.

Monopolar vs. Bipolar

In monopolar mode, electrical current is delivered to the patient via an active cable and electrode. Current returns to the unit through a return electrode pad or plate to disperse the return current, thus preventing focused heat which can cause burns.

In bipolar mode, two electrodes, typically the tips of a pair of forceps or scissors, serve as the equivalent of the active and dispersive leads in the monopolar mode.

Modes of Electrosurgery:

CUT vs COAG

CUT

There are two types of cut modes: blended cut and pure cut. Pure cut is typically used for dissection only. In pure cut mode, the surgeon achieves a cut that is very similar to an incision produced by a medical scalpel. The cut is narrow, deep and the surgeon has little or no control over bleeding. This is the effect obtained by high frequency and low voltage.

In blended cut mode, the surgeon achieves a much wider incision by heating up the tissue and letting it cool. This is achieved by lower frequency and higher voltage than pure cut.

COAG

Coagulation is performed by using high voltage and low frequency. In COAG mode, heat is incapable of explosive vaporization, therefore, resulting in a thermal coagulum, also known as a clot. In COAG mode, the surgeon has more control over bleeding because the tissue is allowed time to cauterize in between contact.

10 Best Practices for ESU Performance Testing

1. Always refer to the manufacturer’s service manual

Manufacturer recommended test procedures should always be followed. Refer to the service manual for performance inspection tasks specific to the device. These service manuals typically recommend an inspection frequency. Complete the performance inspection per the manufacturer’s procedure.

2. Adopt a consistent inspection frequency

If the service manual and inspection procedure from the manufacturer is not available, the frequency of inspection must still be determined. One method for determining how often a medical device should be tested is a risk-based method used by the University of VT Biomedical Engineering Department. This method is described in “Medical Equipment Quality Assurance: Inspection Program Development and Procedures” by J. Tobey Clark et al. available from Fluke Biomedical. For electrosurgery generators this risk-based method yields a twice per year (every 6 months) inspection frequency.

Additionally, most major electrosurgical device manufacturers recommend semi-annual preventive maintenance testing to ensure the performance of the unit.

3. Adopt a formal standardized test procedure

If the service manual and inspection procedure from the manufacturer is not available, it is still the responsibility of the medical facility to choose and standardize a test procedure. It is important that the electrosurgery generator functionality be quantitatively evaluated by comparing it to the manufacturer’s specifications, or the requirements in the applicable medical device standard. When the medical device manufacturer’s specifications are not known, the IEC standard requirements are a reasonable substitute. Once the inspection criteria have been agreed upon, no changes should be made without a rationale statement describing why a change was required, what the change is, and how this change was validated.

4. Pair additional test equipment with your electrosurgical analyzer for comprehensive testing

Most manufacturer performance inspection procedures require electrical safety tests including ground wire resistance and chassis leakage. Keep an electrical safety analyzer close by to complete the electrical safety portion of the performance inspection easily.

Additionally, a medical oscilloscope can be used to show the actual wave shape of the device under test (DUT). This waveform output can be compared to the DUT’s service manual.

5. Be mindful of test leads while testing

Keep all test leads and interconnecting leads as short as possible and do not cross or coil measurement leads. Radio frequency energy behaves differently than low frequency energy. It radiates and induces electrical current flow in addition to any conductive current flow through test leads that cross and coiled leads. When leads are too long they act more like antenna than test leads.

6. Always exercise caution when dealing with active electrodes

Active electrodes present many dangers. Do not touch the active electrode or return pad/plate of the ESU while it is activated in either cut or coagulation mode. Turn off the ESU before adjusting or removing connections.

Additionally, be aware of other flammable hazards:

• Alcohol

• Oxygen

• Moisture

7. Perform all tests necessary to ensure performance

Power distribution/output tests: Power distribution/output tests measure the power output properties of the ESU and supply output current (A), power (W), peak-to-peak voltage (V), and crest factor values. The power distribution test evaluates the output across multiple loads to determine how well the impedance-sensing-circuits of new generation electrosurgery generators automatically adjust the output of the ESU so that it is not reduced by the presented load.

High Frequency (RF) leakage current tests: The RF leakage current in electrosurgical units is a critical parameter to measure because it may cause accidental burns in patients. The particular standard for electrosurgical units, IEC 60601-2-2, indicates the maximum RF leakage levels and defines the elements and their layout to do these measurements.

RECM tests: The RECM (return electrode current monitor) is the “watch-dog” that alarms (both audibly and visually) and prevents the electrosurgery generator from energizing when the high limit threshold for current flowing through the return electrode plate or pad has been exceeded.

Inert gas flow and pressure parameters: In some electrosurgery generators a special option allows an inert gas envelope to be produced encasing the surgery site so that oxygen is eliminated at that specific spot. Oxygen causes charring of the tissue at the site of the surgery. Eliminating oxygen prevents this charring and produces cleaner, more precise incisions. These more precise wounds heal faster, with less opportunity for infection of the tissue. Test the gas flow and pressure for such inert gas outputs.

8. Use test automation to quickly perform tests, document measurements, and archive data

One of the best ways to shorten learning curves for infrequently used test instruments and new or infrequently scheduled testing is to standardize the procedure. Additionally, make sure that pictures accompany the work instructions to show how it looks when test leads are properly placed, where mechanical adjustment and lubrication points are located, and wave form shapes are exemplified.

To ensure that the flow and sequence of the tests are complied with, prevention of doing test out of order ensures that data is consistent and statistically relevant. There should be just one test report in which all test results are documented. Test automation software provides these benefits when coupled with electrosurgical analyzers.

9. Always archive test results

The purpose of testing and producing test results is to have a continuing stream of data showing all changes in the performance and safety of the electro surgery generator year over year.

Long-term trending of this statistically relevant information provides the basis for predictive maintenance (i.e., when the next repair is most likely to happen) so that parts (especially long-lead time and costly parts) can be ordered and received just-in-time for the repair event. This saves money and increases up-time (the amount of time the medical device is available for use).

The best place to archive test result information is in a database/CMMS (computerized maintenance management system). Archiving paper into filing cabinets rarely results in anyone capturing or understanding the long-term implications of failures.

10. Choose to test with an analyzer that you can depend on for complete preventive maintenance and safety testing.

The QA-ES Electrosurgery Analyzer is an ESU tester you can depend on for preventive maintenance and ESU safety testing to test all critical functions.

The Ansur-Automated QA-ES analyzer streamlines ESU preventive maintenance and allows users to increase productivity with the following functions:

• Step-by-step test templates includes pictures, diagrams and hyperlinks

• User-friendly Ansur test automation functionalities

• Checklists and user messages

• Output power, current, peak-to-peak voltage, and crest factor measurements

• Low frequency leakage currents included when using an automation compatible with Fluke Biomedical electrical safety analyzers.

上一篇:没有了

下一篇:Training

Copyright 2008-2021 www.promed-tech.com Promed Technology Co., Limited XMLMAP PROMED

E-mail Contact Whatsapp

Tel

+86-755-33130798

Scan me