Tel
+86-755-33130798
PRODUCTS
Introduction:Features 1. Three parameters:SPO2, Pulse Rate, and TEMP 2. Spo2 waveform(PLETH),PR strength display 3. 2.8inch color TFT LCD in real time display, displayable in big font and big screen 4. Unique software for telemonitoring by transmitti...
Technical Specifications |
1. General |
Monitoring Parameters: SpO2,PR,PLETH,Pulse strength,TEMP |
Signal sockets:SpO2 socket,PR socket,TEMP socket,USB socket(for transmit data and AC power) |
2. Display screen |
Type: 240 × 320 dot matrix color TFT LCD screen |
Display area: 43mm×58mm |
3. Functions |
PLETH: The oximeter can display PLETH waveform |
Audio and visual alarm: The speaker sounds the alarm and the parameter’s measurement will become in red to flash on the screen. Also the same when sensor-off. |
Prompting: Icons of battery capacity, pulse sound switch and alarm sound switch. |
Power saving: The oximeter will automatically shut down to save power. The time is settable. |
Trend: Storage/Review trend data up to 3888 groups, data save time being set freely |
Data storage: The saved data will never be lost even if the oximeter is powered off. |
Communication: Transmitting the patient’s trend to a PC for display, observation, saving and printing via USB or Bluetooth communication |
4. Power Requirements |
Battery type: common 1.5V AA alkaline or rechargeable batteries |
Run time: 6 hours |
5. Environment |
Temperature |
Operation: 0°C ~ 45 °C |
Transportation and storage: -20°C ~ 60 °C |
Humidity |
Operation: 15% - 95 %( noncondensing) |
Transportation and storage: 10% - 95 % (noncondensing) |
Altitude |
Operation: 86 KPa ~106 KPa |
Transportation and storage: 50KPa~106 KPa |
6. Parameter Specifications |
SpO2 |
Patient: Adult, pediatric,neonate |
Range: 0% - 100% |
Resolution: 1 % |
Accuracy: 70% - 99%: ±2 % |
0% - 69%: Unspecified |
PR |
Range: 0 bpm ~250bpm |
Resolution: 1bpm |
Accuracy: ±2bpm |
TEMP |
Channel: 1 |
Input: Body surface thermal-sensitive resistor temperature sensor |
Range: 0℃~50℃ |
Resolution: 0.1℃ |
Accuracy: 0.2℃ |